Troubleshooting
Troubleshooting Assembly Level Problems
1-28
should be at 3 GHz and ≥0 dBm.
2. If the signal is not present or at the correct level, replace the Frac-N.
3. If the signal is present, replace the A8 Output.
WARNING Don’t re-install the A8 Output with the signal generator powered up. This could
cause serious damage to the signal generator.
402 Switching Filter Test
1. Replace the A8 Output.
403 ALC Mod System Test
1. Replace the A8 Output.
404 Pulse Mod System Test
1. Replace the A8 Output.
405 RF Path Test
1. Replace the A8 Output.