User Manual    California Instruments 
154  CSW Series 
MEISTER CURVE GROUP 
If the Meister curve group is selected, the screen will appear as shown in Figure 8-27.  The 
screen has the following parameters that are unique to the group: 
 
Figure 8-27: IEC 61000-4-13 Meister Curve 
1.  LEVEL  Sets the percentage level of the inter harmonics relative to the 
fundamental. The level is fixed for the entire frequency range defined by FRANGE. To 
change levels, the test state must be off and user class must be selected. 
2.  FRANGE  The range is selected by rotating the shuttle.  The range selection is 
required to change the level.  An individual range step can be run by selecting the 
desired range, followed by RUN RANGE from the RUN field. 
3.  STEP  The step size defines the inter harmonics sweep points.  This step size is 
fixed for the entire range of the sweep.  The STEP size can be changed when the test 
state is off. 
4.  PAUSE  Sets the time in seconds for which the test will pause between ranges.  
There is only one value for the entire test.  The PAUSE can be set when the test is not 
running. For the Meister curve test, the pause time is normally set to 0 secs. If a value 
other than 0 is entered, the inter harmonics are set to 0% during the pause times. 
5.  RESONANT  This field is used to report the resonant points, if any, after running the 
Meister curve test.  No editing is allowed in this field.  When the test is completed, a 
display of harmonics current versus frequency plot is available. To view the graph, 
select the field and press the ENTER key. 
8.3.3.3  RUNNING THE TEST 
The test sequence used is a function of the EUT class. The end user needs to determine the 
appropriate class of EUT. The test protocol is document in the IEC 61000-4-13 test standard. 
For reference, the test flow charts are shown here. It is assumed that the end user has a copy of 
the actual test standard however.