User Manual    California Instruments 
70  CSW Series 
  A right arrow indicates the waveform is presently selected for 
the phase. If the cursor is moved to this field, the ENTER key 
will execute the selected MODE. If the mode is set to PROG, 
pressing ENTER while the cursor is on the user defined entry 
will select the custom waveform for the phase shown in the top 
right corner of the display. 
  If the MODE is set to either VIEW option, the waveform data 
under the cursor will be displayed when the ENTER key is 
pressed. Press the ENTER key again to return to the 
WAVEFORMS menu. 
3.11.8  ADVANCE MEAS. Menu   
This entry in the MENU 2 screen displays the HARMONICS/TRACE ANALYSIS screen which is 
covered in section 4.6. This field can be used in lieu of the MEAS key to directly bring up the 
advanced measurements screens. 
3.11.9 APPLICATIONS Menu 
 
Figure 3-24:  APPLICATIONS menu 
The APPLICATIONS menu provides access to the optional application specific pre-programmed 
test sequences. Since these test sequences are optional, this menu may have no choices if 
none of the options are installed. The following entries may be found in the APPLICATIONS 
menu: 
Entry Description 
 
MIL-STD 704  Test sequence for MIL standard 704 AC and DC tests. 
RTCA/DO-160  Test sequence for RTCA DO160 commercial aviation AC and 
DC tests. 
IEC-1000-4-11  Test sequences for IEC 1000-4-11 Voltage Dips and Variations 
test standard. 
IEC-1000-4-13  Test sequence for IEC 1000-4-13 Harmonics and Inter 
harmonics test standard. 
WH METER  Watt Hour meter measurement option. Tracks energy usage 
over a period of time and calculates Watt Hours used.