12 2014-07-30 127139 13
toxic fumes. Proper ventilation must be used.
To avoid hazards from treating fluids in a wrong manner,
always reduce the "Max. SET-temperature allowed” in
the CALIBRATOR SETUP MENU according to the
specifications of the fluid to be used.
If using a calibrator outside of the fluids specifications
there is a risk of fire hazards, personal Injury or chemical
release.
By reducing the "Max. SET-temperature allowed”, the
calibrator cannot be used outside this temperature
range.
Be aware of the flash point, the boilling point and other
fluid properties applicable to the usage when setting the
Max. SET-temperature. Read the MSDS (Material
Safety Data Sheet) of the liquid before use.
Always remove the liquid from the calibrator before
transportation.
Product information on the fluid must be carefully
investigated before use.
Do not handle hot fluid.
If the oil is heated beyond the flash point, it may
constitute a fire hazard.
Do not pour water or any other fluids into a bath filled
with hot oil, because only a few drops of water might
cause a steam explosion, if poured into above 100°C
hot oil.
Do not under any circumstances pour water on burning
oil. It might cause a dangerous steam explosion.
Caution – Hot surface
This symbol is engraved in the grid plate.
Do not touch the grid plate, the well or the insertion
tube when the calibrator is heating up – they may be
very hot and cause burns.
Do not touch the lid or the spill tray when the calibrator
is heating up – they may be very hot and cause burns
(RTC-158/250 A/B/C only).