127139 13 2014-07-30 155
INPUT SPECIFICATIONS
ACCURACY IN C/F
Accuracy thermocouple type E input
-200°C to 1000°C
(excluding sensor accuracy)
±0.18°C(±0.32°F) @ -200°C(-328°F)
±0.09°C(±0.17°F) @ -50°C(-58°F)
±0.06°C(±0.11°F) @ 0°C(32°F)
±0.06°C(±0.11°F) @ 155°C(311°F)
±0.07°C(±0.12°F) @ 320°C(608°F)
±0.08°C(±0.14°F) @ 650°C(1202°F)
±0.10°C(±0.29°F) @ 1000°C(1832°F)
Accuracy thermocouple type J input
-210°C to 1200°C
(excluding sensor accuracy)
±0.23°C(±0.41°F) @ -210°C(-346°F)
±0.10°C(±0.18°F) @ -100°C(-148°F)
±0.08°C(±0.14°F) @ 0°C(32°F)
±0.09°C(±0.16°F) @ 155°C(311°F)
±0.09°C(±0.16°F) @ 320°C(608°F)
±0.09°C(±0.17°F) @ 650°C(1202°F)
±0.13°C(±0.23°F) @ 1200°C(2192°F)
Accuracy thermocouple type K input
-200°C to 1372°C
(excluding sensor accuracy)
±0.27°C(±0.49°F) @ -200°C(-328°F)
±0.14°C(±0.24°F) @ -100°C(-148°F)
±0.10°C(±0.19°F) @ 0°C(32°F)
±0.11°C(±0.20°F) @ 155°C(311°F)
±0.11°C(±0.20°F) @ 320°C(608°F)
±0.13°C(±0.23°F) @ 650°C(1202°F)
±0.20°C(±0.36°F) @ 1372°C(2502°F)