86 2014-07-30 127139 13
Min SET temp / Max SET temp:
Enter the access code to get access to the editor.
Use the NUMERIC keys to set the Min/Max SET
temperature in Celsius, Fahrenheit or Kelvin.
Note…
The Enter Access Code box is displayed every time you try
to access the Min/Max SET temp parameters. Type in your
access code and continue.
Access code:
T
he following features can be protected by an access code:
Resetting the calibrator to Factory default settings.
Setting the Min/Max SET Temperature.
Editing the Access code while it is enabled.
Press or to access the Access code function.
Use the NUMERIC keys to type in a value from 0000 to
9999. Use all 4 digits.
Typing 0000 disables the Access code function.
The access code is accepted showing a green check for
a few seconds allowing you to continue.