6.4.9 Output Calibration Summary
The following Table is a summary of the preceding calibration steps. Note that Series II MX
models have fewer calibration coefficients.
Program the following values in the table and make the adjustments in the OUTPUT
CALIBRATION screen. Select the phase to be calibrated by pressing the PHASE key.
150 VAC range Volt Hi Freq
Repeat the adjustments at 60 and 400 Hz until the output is within ± 0.05 volts
300 VAC range, 0.0 V,
60 Hz
300 VAC range Volt Hi Freq
Repeat the adjustments at 60 and 400 Hz until the output is within r 0.05 volts
150 VAC range, 120V,
60 Hz
R9, R10, R11
INTER HARM FS
20 ± 1.0 Vrms
20 ± 1.0 Vrms
Table 6-3: Output Calibration Table – MX Series I
300 VAC range, 0.0 V,
60 Hz
300 VAC range High Freq
DC Zero, -HF models only.
300 VAC range, 0.0 V,
820 Hz
150 VAC range, 120V,
60 Hz
400 Hz, 20VAC
1800 Hz, 20VAC
R9, R10, R11
INTER HARM FS
20 ± 1.0 Vrms
20 ± 1.0 Vrms
Table 6-4: Output Calibration Table – MX Series II