with mode of operation such as
programming frequency when in
DC mode.
Command issued but unit
is not in remote state
Put instrument in remote state
before issuing GPIB
commands.
Some commands are supported
by the unit but are locked out
for protection of settings and
are not user accessible.
Problem with trigger
system.
Unit could not generate trigger
for transient execution or
measurement.
Trigger request has been
ignored.
Trigger setup incorrect or unit
was not
armed when trigger
was received. Check transient
system or measurement trigger
system settings.
Initialization request has
been ignored
Unit was told to go to armed
state but was unable to do so.
Could be caused by incorrect
transien
t system or
measurement acquisition setup.
Incorrect parameter or
parameter value. Check
programming manual for
allowable parameters
Requested setting conflicts
with other setting in effect.
Check other settings. E.g. trying
to program a DC offset while in
AC mode
Parameter data outside of
allowable range.
Check programming manual for
allowable parameter values
More data received than
expected
Check programming manual for
number of parameters or data
block size
"Illegal parameter
value"
Parameter value is not
supported
Check programming manual for
correct parameters
One or more transient lists
programmed has different
length.
All lists must be of same length
or transient cannot be compiled
and executed.
No storage space left to
save settings or data.
Delete other settings or data to
make room.
Too many waveform
directory entries
Delete one or more waveforms
from waveform memory to
make room.
Waveform requested not in
directory
Check waveform directory for
waveform names present.
Too many or non ASCII
characters used in waveform
file definition.
Use ASCII characters only
Check hardware for proper