User Manual - Rev AV AMETEK Programmable Power
MX Series 252
Number
Message String
Cause
Remedy
not be recalled.
overwrite old content.
6
"Limit memory lost"
Hardware configuration
settings lost.
Contact CI service department
at
support@calinst.com
to
obtain instructions on restoring
configuration data.
7
"System memory lost"
Memory corrupted.
Recycle power.
8
"Calibration memory
lost"
Calibration data lost.
Contact CI service department
at
support@calinst.com
to
obtain instructions on restoring
calibration data or recalibrate
unit.
9
"Start angle must be
first sequence"
Start phase angle in wrong
place..
Start phase angles can only
programmed at the start of a
transient list. Once a transient is
in progress, phase angle cannot
be changed.
10
"Illegal for DC"
Operation not possible in
DC mode.
Switch to AC or AC+DC mode.
11
"Duplicate sequence"
Transient list sequence
number already used.
User new or available sequence
number instead.
12
"Too many sequence"
Number of transient list
steps exceeds maximum.
Reduce the number of steps in
the transient list. (Max = 32 for
Series I or 100 for Series II).
13
"Missing list
parameter"
One or more transient list
parameters missing.
Check programmed lists.
14
"Voltage peak error "
Peak voltage exceeds
internal bus voltage
This error may occur when
selecting user defined wave
shapes with higher crest
factors. Reduce programmed
RMS value.
15
"Slew time exceed
dwell"
Time needed to slew to
final value is less than
dwell time.
Check dwell times in transient
list settings. Increase dwell time
or change slew rate for affected
parameter.
16
"Illegal during
transient"
Operation requested not
available while transient is
running.
Wait till transient execution is
completed or abort transient
execution first.
17
"Output relay must be
closed"
Operation not possible with
open relay
Close relay before attempting
operation. E.g. transient
execution requires output relay
to be closed.
18
"Trans. duration less
then 1msec"
Dwell time below minimum
or 1 msec
Increase dwell time to at least 1
msec.
19
"Clock and sync must
be internal"
Operation not possible with
external clock
Switch to internal sync.
(Default)
20
"Input buffer full"
Too much data received.
Break up data in smaller blocks.
21
"EOS Fault"
Hardware error reported by
EOS option.
This option is not available
on MX Series products.
Cycle power on EOS to reset
error. If error persists, contact
CI service at
support@calinst.com
for
repair.
22
"Waveform harmonics
limit"
Harmonic contents of user
defined wave shape are
Reduce harmonic content or
reduce fundamental frequency