DATA MODE  Selects absolute or relative harmonics display for TABLE and 
BAR view modes. In relative mode, all harmonics are shown in a 
percentage of the fundamental, which is normalized at 100 %. In 
absolute mode, the harmonic amplitudes are shown in absolute 
volts or amperes. 
  This mode does not apply to the TRACE view display mode and 
is ignored when this mode is selected. 
SCALE  Sets the horizontal time axis for the TRACE view display mode. 
The field range is: 
Series I:  4 ms to 104 ms in single-phase mode (MX30-3Pi or 
MX45-3Pi only) or 12 ms to 312 ms in three phase mode. 
Series II:  4 ms to 42 ms in single-phase mode (MX30-3Pi or 
MX45-3Pi only) or 12 ms to 128 ms in three phase mode. 
  This parameter is ignored when the TABLE or BAR view display 
mode is selected. 
TRIG MODE  This field sets the trigger mode for the acquisition. Available 
options are SINGLE (single shot acquisition) or CONT 
(continuous acquisition). In SINGLE shot mode, the acquisition 
is triggered once each time the START field is selected and the 
ENTER key is pressed. The selected trigger source is used to 
determine the trigger point. Once the acquisition has been 
triggered, the data are displayed and do not change until the 
next acquisition is triggered. This mode is most appropriate for 
single shot events such as start up currents. 
  In the CONT mode, acquisitions occur repeatedly and the data 
is updated on screen after each trigger occurs. This provides a 
continuous update of the data and is most appropriate for 
repetitive signals. 
TRIG SOURCE  The trigger source selects the event that will trigger a 
measurement acquisition. Available options for this field are 
IMM (immediate), PHASE A or SET VOLT. The IMM trigger 
source causes the acquisition to trigger immediately when the 
ENTER key is pressed on the START field. Essentially, this is 
an asynchronous trigger event. The acquisition will always be 
triggered in this mode and data is available immediately. 
  The PHASE A source will cause the acquisition to trigger on the 
occurrence of a set phase angle for the voltage on phase A. The 
trigger source is always phase A when in this mode, regardless 
of the phase selection shown in the top right corner of the 
display. When the acquisition is started, the acquisition system 
waits for the specified phase angle to occur before triggering the 
acquisition. This mode allows exact positioning of the acquisition 
data window with respect to the voltage waveform. 
  The SET VOLT mode causes the acquisition to trigger at the 
specified voltage. This mode also programs the selected phase 
or all three phases when the measurement is started from the 
START field. As such, this trigger source selection also 
programs the output voltage to the selected rms level. 
TRIG PHASE / SET VOLT  This field changes purpose, depending on the trigger source 
selected immediately above it. If the trigger source equals IMM 
or PHASE A, this field can be used to program the trigger phase