Quality Control Testing
The remaining lockout time together with the date is displayed in the information
windows of the Main screen.
The maximum displayed negative value is –90. It may mean that more than 90
days has passed since the limit or a successful QC was never performed.
9.3 Recall QC Results
All QC measurements are stored in the QC memory, which is separated from the
memory for the patients’ measurements. The Urilyzer 100 has memory for 500
(PRO 1000) QC measurements.
See
8. Recall Results for more information on how to recall and view results
from a database.
Only the additional information specific to the QC database is described in this
chapter.
In the list view the good results have black text, while any failed results have red
text.
In the QC result screen PASSED text is displayed after the good QC result Id,
while red FAILED text is displayed after the failed QC result Id. For failed QC solu-
tion results the out of range pad results are also marked in red.
The QC solution results have a second page, where the set limits are listed.