13. Troubleshooting
Your Urilyzer 100 analyzer will operate properly if you follow the directions for
using and cleaning the instrument.
Advisory messages will be displayed when your attention is required on any
disorder or result of a performed action.
The user interface messages can be categorized into the following groups:
1. Error messages
2. Warning messages
3. Information messages
The active errors and warnings can be listed by pressing the status bar area at
any screen.
Error messages
If an error prevents the instrument from being used, certain selection areas on
the screen will be disabled and testing procedure cannot be started. The back-
ground of the status bar changes to red. Performing the displayed corrective
action will remove the error and allow you to use the instrument and enable
testing.
Warning messages
Errors of less importance are categorized into the warning messages group.
These kinds of errors do not prohibit testing, but may limit certain functionality
(i.e. transfer, printing) of the system. The background of the status bar changes
to orange. These errors do not compromise the testing and the measurement
performance of the system. The resolution of these errors may include the restart
of the system. When you have taken the corrective action, the message will be
removed from the system.
Information messages
Provides a feedback about the successful execution of an action and/or provides
additional information for the operator.
Based on the presentation the display type of the messages are the following:
1. Status line: appears permanently in the status bar
2. Timed pop-up window: a pop-up window is displayed only for a few seconds,
then disappears automatically without operator interaction
3. Pop-up window: a pop-up window requiring confirmation by the operator to
clear up disappears after the confirmation of the operator
4. Result view: message appears on the standard content area