48
49
Options menu
NOTE: These features can be modified by any operator and stored separately for
each operator (Pro)
Fast mode (serial reading): if enabled, the test strip is measured directly after
Measure is pressed on the Measurement screen (note: in fast mode the large
start button is renamed to Measure and the background is changed to orange).
In this case, it is up to the user to time the incubation period outside the analyzer.
When working in Fast Mode, ensure that you have a foolproof system for match-
ing sequence numbers to samples.
The status of fast mode cannot be saved. After logout or system restart the
analyzer always starts in Normal mode.
When performing serial measurements in Fast Mode, allow the strips
to react for approximately 60 seconds before inserting them in the
analyzer and pressing MEASURE. False-low or false-negative results
may be obtained for some parameters if the reaction time is too short.
Likewise, false-high results may be obtained for some parameters if
the incubation time outside the analyzer is too long.
The option to enable Fast mode only appears on the User options screen, if this
option is enabled on the Settings»Measurement screen.
Sound: if enabled the analyzer confirms the touching actions with a short beep
sound
LCD brightness: Use the left and right buttons to change the brightness of the
LCD display or click on the input field to set the LCD brightness value from a
numeric keyboard.
Change passw.: (Pro only) the active operator may change the password by
pressing the
Change passw. button. First the system asks for the cur-
rent password, than the new password has to be repeated twice. The system
confirms the successful change.
The change passw. button appears only if an operator with a password is
logged in to the system. In case of ‘autologin’ operator this button does not
appear.
The minimum password length is 3 characters.