68
Troubleshooting
69
Legend
Categories (C.): Type (T.):
1. E Error messages 1. S Status line
2. W Warning messages 2. TP Timed pop-up window
3. I Information messages 3. P Pop-up window
4. R Result view
ID C. T. Text Long text Action
E99
E S Head
HW
Head hardware error.
Please call Service.
Contact your service repre-
sentative
E98
E S Printer
HW
Printer hardware error.
Please call Service.
Contact your service repre-
sentative.
E97
E S Head
voltage
Head voltage value is out of
range. Please call Service.
Contact your service repre-
sentative.
E96
E S Power
voltage
Power voltage value is
out of range. Please call
Service.
Contact your service repre-
sentative.
E90
E S Refer-
ence pad
Failure of reference pad
check. Reference pad value
of the tray is out of range.
See User’s Manual for
further instructions.
Reference pad is contami-
nated or damaged.
Clean the test strip tray and its
reference pad (Chapter 12.2).
If the error remains, replace
the test strip tray with a spare
one or request a new one.
If the error remains, Contact
your service representative.
E89
E S QC lock-
out
Go to „QC measurement”
to perform QC check.
Perform QC check measure-
ments to remove the QC
lockout.
E88
E S Memory
limit
Database limit exceeded,
please delete results to
free up space.
Free up memory by erasing
old data!
W69
W S Output
port
Output port not open.
Please restart the system!
Restart analyzer.
W68
W S Output
internal
Output internal error.
Please restart the system!
Restart analyzer.
W67
W S Output
init
Output not inited.
Please restart the system!
Restart analyzer.
W66
W S Output
closed
Output closed.
Please restart the system!
Restart analyzer.
W65
W S Output
memory
Not enough memory for
output.
Please restart the system!
Restart analyzer.