8.5. Simple Comparison Loss-in-weight Batching
When the weight value is compared with the setpoint and the condition is met, the control output turns ON.
Condition Control output
Weight value ≤ Near-zero
Near-zero
Gross value ≥ Full
Full
Net value ≥ Final
Preliminary Medium flow
Net value ≥ Final
Free fall Small flow
Net value < Final
Underweight
Underweight
Net value > Final
Overweight
Overweight
Final
Overweight ≥
Net value ≥ Final
Underweight OK
Timing chart
8.6. Simple Comparison Functions
●Automatic print
When the weight value is five times the minimum division or more and the large flow, medium flow and small
flow outputs turn ON, data is output once from the standard serial output or options. To output again, wait for
the –weight value to be four times the minimum division or less.
Final
all