21.11. Setpoint Input Functions
SP F Modbus Item Setting Value Default
User
setting
01
401001
401002
Setpoint setting method
0: Setpoint input not used
1: 5-digit input from external devices
2: 16-digit input from external devices
0
21.12. Analog 4-20 mA Output Functions
An F Modbus Item Setting Value Default
User
setting
01
401101
401102
Output weight value
1: Displayed weight value
2: Gross value
3: Net value
1
02
401103
401104
Weight value at 4 mA
output
-999999 to 9999999
0
03
401105
401106
Weight value at 20 mA
output
-999999 to 9999999
20000
21.13. Modbus TCP Functions
EthF Modbus Item Setting Value Default User setting
IP address
XXX.
XXX.
XXX.
XXX
0.0.0.0
to
255.255.255.255
192.168.0.2
01
-
02
-
03
-
04
-
Subnet 255.
255.
XXX.
0
Available only to
255.255.0.0
or
255.255.255.0
255.255.255.0
mask
05
-
06
-
07
-
08
-
Default XXX.
XXX.
XXX.
XXX
0.0.0.0
to
255.255.255.255
0.0.0.0
gateway
09
-
10
-
11
-
12
-