EasyManua.ls Logo

AND AD-4401A - Format When Communication Type (Rs F-02) Is 5 or 6

AND AD-4401A
100 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
AD-4401A
Page 52
14.7. Format When Communication Type (rS F-02) is 5 or 6
Data format (Comparison result + Weight value)
Comparison
result
Gross value/Net value/
Output weight value
Terminator
2 chars 6 chars 2 chars/1 char
Item
Value
Comparison result
F F
*1
Gross value (rS F-02=5)
Net value (rS F-02=6)
Output weight value (rS F-01)
0 0 0 0 7 F *2
*1: Comparison results are converted from 8 binary digits into 2 hexadecimal digits.
2 hexadecimal digits
16
1
16
0
8 binary digits 0: OFF 1: ON
2
7
2
6
2
5
2
4
2
3
2
2
2
1
2
0
Near-zero
Underweight/Hi-Hi
Overweight/Hi
Large flow/Go
Medium flow/Lo
Small flow/Lo-Lo
Discharge
Weighing end
*2: The gross value and the net value are converted from a signed decimal number into a 6-digit
hexadecimal number. Negative numbers are represented by the two's complement. A decimal point is
not included.
Decimal
Hexadecimal
+127
00007F
2
000002
1
000001
0
000000
-1
FFFFFF
-2
FFFFFE
-127
FFFF81

Table of Contents