Interference Analyzer (Option 25) 3-17 Sweep Menu
Spectrum Analyzer MG PN: 10580-00349 Rev. H 3-75
Sweep Mode Menu
Key Sequence: Shift > Sweep (3) key > Sweep Mode
Sweep Mode is not applicable in Zero Span.
Fast: Fastest sweep speed (up to 100 times faster than Performance).
Performance: Provides best amplitude accuracy and ensures all
specifications are met.
No FFT: Slowest sweep speed. Suited for analog and pulse modulated
signals.
Burst Detect: Press to view narrow pulsed signals. This function is
available on MS2720T and MT8220T models only.
Show Help: Displays a table showing the merits and trade-offs of the
Sweep Mode settings.
Back: Returns to the “Sweep Menu” on page 3-74.
Figure 3-69. IA Sweep Mode Menu
Sweep Mode
Back
Performance
No FFT
Show Help
Fast
Burst Detect