5-2 Procedure CW Signal Generator (Option 28)
5-4 PN: 10580-00349 Rev. H Spectrum Analyzer MG
4. Press the Amplitude key and set the power level to High or Low. The typical nominal
output power in the high setting is about 0 dBm. The typical nominal output power in
the low setting is about –30 dBm.
5. Change the settings on the attenuator to adjust the power level. The large knob changes
the power in 10 dB steps and the small knob adjusts the power level in 1 dB steps.
6. Press the Offset submenu key to add an offset (in dB) to the amplitude level. This offset
compensates for any attenuation that is placed in-line between the splitter and the
DUT. Offset range is +100 dB to –100 dB.
Figure 5-3. CW Signal Generator Frequency Menu