3-3 General Measurement Setups W-CDMA/HSPA+ Signal Analyzer
3-6 PN: 10580-00234 Rev. T 3GPP MG
Threshold Setup
The threshold level is an advanced setting that can be set to indicate which codes are
considered active. In the Code Domain Power screen, the threshold level is indicated by a
horizontal dotted red line. Any code channels exceeding this power level are considered active
traffic channels and any code channels below this power level are considered inactive or noise.
To set the threshold level:
1. Press the Setup main menu key, then press More.
2. Press the Threshold submenu key and select Auto or Manual.
3. To set the threshold level manually, press the Manual Threshold submenu key and use
the rotary knob, arrow keys, or the numeric keypad to change the value. When entering
a threshold using the keypad, the screen menu will show the Units key. Press the dB
submenu key or the Enter key to set the entered threshold.
Filtered versus Unfiltered Power
The ACLR measurement uses the filtered channel power to determine the ACLR values and
it is listed as filtered on the display. In all other screens the unfiltered channel power is
displayed as channel power.
ООО "Техэнком" Контрольно-измерительные приборы и оборудование www.tehencom.com

Table of Contents

Other manuals for Anritsu Spectrum Master MS2720T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Anritsu Spectrum Master MS2720T and is the answer not in the manual?

Anritsu Spectrum Master MS2720T Specifications

General IconGeneral
BrandAnritsu
ModelSpectrum Master MS2720T
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals