Applied Biosystems QuantStudio 7 Flex
194 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #102 background imageLoading...
Page #102 background image
3.
Open the access door.
4.
Remove the lamp from the system:
a.
Slide the lamp release lever downward.
b.
Firmly grasp the lamp and lift it up and out of the sloed mount.
5.
Inspect the lamp for damage (carbon usually coats the inside of a failed lamp). If
necessary
, dispose of the lamp and obtain a replacement.
IMPORTANT! Dispose of the lamp in accordance with your local municipal
waste ordinances. Do not dispose of the lamp as unsorted municipal waste.
6.
Install the new lamp into the QuantStudio
6 or 7 Flex Real-Time PCR System:
a.
Slide the lamp release lev
er upward.
b.
Firmly grasp the lamp, place it into the sloed mount, then carefully slide
the lamp downward into place.
Appendix B Maintenance
Replace the halogen lamp
B
102
QuantStudio
6 and 7 Flex Real-Time PCR Systems v1.6 Maintenance and Administration Guide