Guidelines for preparing calibration plates and array cards
Observe the following guidelines when preparing calibration plates and array cards:
Wear appropriate protective eyewear, clothing, and gloves.
Prepare and run calibration plates and array cards within the recommended time
limits listed below.
Consumable Time to thaw
After thawed,
run within…
All 96-/384-well calibration consumables
[1]
30 minutes 120 minutes
All array card calibration consumabes
[2]
30 minutes 120 minutes
RNase P plate (96-/384-well) 5 minutes 30 minutes
RNase P array card
[3]
15 minutes 60 minutes
[1]
A
ll 96-well plates and 384-well plates used in the calibration of the instrument (ROI, background,
uniformity, and dye).
[2]
All array cards used in the calibration of the QuantStudio
7 Flex System only (ROI, background,
uniformity, dye, and normalization).
[3]
For use with the QuantStudio
7 Flex System only.
(Plates only) Do not discard the packaging for the calibration plates. Each plate
can be used to calibrate the QuantStudio
6 and 7 Flex Systems 3 times for up to
6 months if it is stored in its packing sleeve.
Chapter 2 Calibr
ate the instrument
Guidelines for preparing calibration plates and array cards
2
36
QuantStudio
6 and 7 Flex Real-Time PCR Systems v1.6 Maintenance and Administration Guide