Note: Conrm that the replacement heated cov
er supports the plate or array card
format that you want to use. Some heated covers support more than one consumable
type.
1.
Power o and unplug the QuantStudio
™
6 or 7 Flex Real-Time PCR System, then
allow it to cool for 15 minutes.
2.
Open the access door.
3.
Unlock the heated cover by pinching the handle together, then pull the assembly
from the system and place it on a clean, dry surface.
4.
Install the new heated cov
er into the QuantStudio
™
6 or 7 Flex Real-Time PCR
System.
IMPORTANT! Align the arrows on the front handle as shown below. If the
arrows are not aligned, push the heated cover into the system until the handle
locks into place.
Correct Incorrect
5.
If y
ou are installing a heated cover of a dierent consumable format, replace the
sample block and plate adapter if necessary.
Change the heated
cover
Appendix B Maint
enance
Change the heated cover
B
110
QuantStudio
™
6 and 7 Flex Real-Time PCR Systems v1.6 Maintenance and Administration Guide