b.
In the Instrument Console, select your instrument from the list of
instruments on the network, then click Add to My Instruments.
Note: You must add a system to your list before you can manage it.
c.
After the system is added to your list, select it, then click Manage
Instrument.
d.
In the Instrument Manager, click
Maintenance, then click Calibration
Reminders.
2.
In the Calibration Reminders Seing table, congure the notication seings for
the calibrations in interest. For each calibration that y
ou want to monitor:
a.
In the Expiry Interval column, enter the number of days that elapse before
the type of calibration expires on the system.
b.
In the Send a Reminder column, select the check box to congure the
QuantStudio
™
Real-Time PCR Software to email a reminder to perform the
calibration.
c.
In the Reminder Interval column, enter the number of days that elapse
before the software emails recipients a reminder to perform the calibration.
3.
In the Enter e-mail addresses for notications eld, enter the email address(es)
that you want to receive email notications. Separate multiple email addresses
with commas (
,
).
4.
Click Apply to change the notication seings.
Appendix C Netw
orking
Monitor the QuantStudio
™
6 or 7 Flex Real-Time PCR System
C
122
QuantStudio
™
6 and 7 Flex Real-Time PCR Systems v1.6 Maintenance and Administration Guide