19. Ap Hi-Lmt (PROGRAM Mode only) - gives the operator the flexibility to
program a maximum peak current threshold or ceiling level. When this level
is reached a failure will occur. Ap Hi-Lmt is only selectable when the source
is configured for AC output.
20. Ap Lo-Lmt (PROGRAM Mode only) - gives the operator the flexibility to
program a minimum peak current threshold or floor level. If a minimum peak
current level is not reached a failure will occur. This ensures a load is
attached to the power source and there is a minimum peak current present.
Ap Lo-Lmt is only selectable when the source is configured for AC output.
21. CF Hi-Lmt (PROGRAM Mode only) - gives the operator the flexibility to
program a maximum crest factor threshold or floor level. When this level is
reach a failure will occur. CF Hi-Lmt is only selectable when the source is
configured for AC output.
22. CF Lo-Lmt (PROGRAM Mode only) - gives the operator the flexibility to
program a minimum crest factor threshold or floor level. If a minimum crest
factor level is not reached a failure will occur. CF Lo-Lmt is only selectable
when the source is configured for AC output.
23. PF Hi-Lmt (PROGRAM Mode only) - gives the operator the flexibility to
program a maximum power factor threshold or ceiling level. When this level
is reached a failure will occur. PF Hi-Lmt is only selectable is only selectable
when the source is configured for AC output.
24. PF Lo-Lmt (PROGRAM Mode only) - gives the operator the flexibility to
program a minimum power factor threshold or floor level. If a minimum power
factor level is not reached a failure will occur. This insures a load is attached
to the power source and there is a minimum power factor present. PF Lo-Lmt
is only selectable when the source is configured for AC output.
25. VA Hi-Lmt (PROGRAM Mode only) - gives the operator the flexibility to
program a maximum apparent power threshold or floor level. When this level
is reached a failure will occur. VA Hi-Lmt is only selectable when the source
is configured for AC output.
26. VA Lo-Lmt (PROGRAM Mode only) - gives the operator the flexibility to
program a minimum apparent power threshold or floor level. If a minimum
apparent power level is not reached a failure will occur. VA Lo-Lmt is only
selectable when the source is configured for AC output.