15
DISPLAY
Touch Current Display (RMS)
Range 1
0.0uA ~ 32.0uA, frequency
DC, 15Hz - 1MHz
Resolution
0.1uA
Accuracy
DC , 15Hz < f <100KHz: ±(2%
of reading + 3counts)
100KHz < f < 1MHZ : ±5% of
reading (10.0uA - 999.9uA)
Range 2
28.0uA ~ 130.0uA, frequency
DC, 15Hz - 1MHz
Resolution
0.1uA
Accuracy
DC , 15Hz < f <100KHz: ±(2%
of reading + 3counts)
100KHz < f < 1MHZ : ±5% of
reading (10.0uA - 999.9uA)
Range 3
120.0uA ~ 550.0uA, frequency
DC, 15Hz - 1MHz
Resolution
0.1uA
Accuracy
DC , 15Hz < f <100KHz: ±(2%
of reading + 3counts)
100KHz < f < 1MHZ : ±5% of
reading (10.0uA - 999.9uA)
Range 4
400uA ~ 2100uA, frequency
DC, 15Hz - 1MHz
Resolution
1uA
Accuracy
DC , 15Hz < f <100KHz: ±(2%
of reading + 3counts)
100KHz < f < 1MHZ : ±5% of
reading (10uA - 8500uA)
Range 5
1800uA ~ 8500uA, frequency
DC, 15Hz - 1MHz
Resolution
1uA
Accuracy
DC , 15Hz < f <100KHz: ±(2%
of reading + 3counts)
100KHz < f < 1MHZ : ±5% of
reading (10uA - 8500uA)
Range 6
8.00mA ~ 20.00mA, frequency
DC, 15Hz 100kHz

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Associated Research 620L and is the answer not in the manual?

Associated Research 620L Specifications

General IconGeneral
BrandAssociated Research
Model620L
CategoryTest Equipment
LanguageEnglish

Related product manuals