42
Save As
To save an edited file without overwriting the original or to give a file a new name,
press the “Save As” soft key in the File Setup screen.
Delete
To delete a file use the up and down arrow keys and scroll the highlighted area to the
file you wish to delete. Press the “Delete” soft key. A warning will appear on the
screen asking if you want to delete the file. Press the ENTER key to complete the
deletion or press the EXIT key to escape.
Load
To load a test file use the up and down arrow keys and scroll the highlighted area to
the file you wish to load. Press the “Load” soft key.
Connect (conditional)
If a file is created with 30 test steps the “Connect” soft key will appear in the File Setup
screen. Press the “Connect” soft key to connect step 30 of one test file to step 1 of
another test file. This feature will only appear if the test file has 30 steps. Only
sequential test files may be linked together (i.e. File 01 File 02).
4.3.6. Fail Stop
Fail Stop is a function that will stop a sequence of tests if a failure occurs. If this
function is turned OFF the sequence of tests will continue to the end of the sequence
regardless of whether or not a failure has occurred. If the Fail Stop is OFF and a
failure occurs during the test sequence, the RESET button will light and a short alarm
will sound but the sequence will continue. At the end of the test sequence, the
RESET button will light and alarm will sound indicating failure during the sequence.
Pressing the RESET button will silence the alarm and reset the instrument.
Turn the Fail Stop function ON and OFF by pressing the “Fail Stop” soft key. Fail Stop
is a parameter that is stored within the test file but may also be turned ON and OFF
temporarily from the Perform Tests screen. If security is enabled you may not turn
Fail Stop ON and OFF using the “Fail Stop” soft key. Fail Stop automatically defaults
to the setting stored in the file when security is activated.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Associated Research 620L and is the answer not in the manual?

Associated Research 620L Specifications

General IconGeneral
BrandAssociated Research
Model620L
CategoryTest Equipment
LanguageEnglish

Related product manuals