86
COMMAND
NAME
VALUE
LS?
List Step Parameters
LS <step number>?
List Step Parameters by step
number
step number = 1-30
SS?
List Selected Step
step number = 1-30
SF?
List Fail Stop condition
1=On, 0=Off
TD?
Read the active data being displayed on the LCD display while the test is in process.
Will also read the last data taken when the test sequence has completed. Each
parameter is separated by commas and includes step number, test type, test status,
and metering. The syntax for this command response is {step, test type, status, meter
1, meter 2, meter 3}. Each meter will contain only the value and not the units.
RD <step number>?
Read the results for an individual step. The step number is the actual step number
that has been saved within the file, not the order of which the steps were executed.
For example if the test was executed starting from step 3 and ending with step 5 then
the first step test results will be found in location 3 not in location 1. Each parameter
is separated by commas and includes step number, test type, test status, and
metering. The syntax for this command response is {step, test type, status, meter
1,meter 2,meter 3}. Each meter will contain only the value and not the units.
RD <file number>, <step number>?
Read the results for an individual step when multiple files are linked. The file number
is the actual location to which the file has been saved. The step number is the actual
step number that has been saved within the file, not the order of which the steps were
executed. For example if the test was executed starting from step 3 and ending with
step 5 then the first step test results will be found in location 3 not in location 1. Each
parameter is separated by commas and includes step number, test type, test status,
and metering. The syntax for this command response is {step, test type, status, meter
1,meter 2,meter 3}. Each meter will contain only the value and not the units.
RR?
Read the remote Reset input signal. When the remote reset has been activated by
closing the contacts the query will return a value of 1 to indicate the instrument is
being Reset.
RI?
Read the remote Interlock input signal. When the remote Interlock has been activated
by opening the contacts the query will return a value of 0 to indicate the instrument is
in the Interlock state and will not be able to generate output voltage or current.
RS?
Read Scanner Status command will respond with a value that identifies the number of
scanners installed or connected to the instrument. Values 0 4 will indicate if there

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Associated Research 620L and is the answer not in the manual?

Associated Research 620L Specifications

General IconGeneral
BrandAssociated Research
Model620L
CategoryTest Equipment
LanguageEnglish

Related product manuals