00H=OK
01H=TEST EEPROM ERROR
Clear Standard Event Status
Register
Clear Service Request Register
Operation Complete Command
When all selected pending
operations complete, ESR BIT0=1
When all selected pending
operations complete, Output
Queue=1
Power-on Status Clear Command
1 = Power-on clear enable
registers
0 = Power-on load previous enable
registers
Power-on Status Clear Query
Standard Event Status Register
Query
Standard Event Status Enable
Command
Standard Event Status Enable
Query
Service Request Enable Command
Service Request Enable Query
*IDN?
Read the instrument identification string. Company = Associated Research Inc.
*RST
Reset the instrument to original power on configuration. Does not clear Enable
register for Standard Summary Status or Standard Event Registers. Does not clear
the output queue. Does not clear the power-on-status-clear flag.
*TST?
Performs a self-test of the instrument data memory. Returns 0 if it is successful or 1 if
the test fails.
*CLS
Clears the Status Byte Summary register and Event registers. Does not clear the
Enable registers.