Chapter 3 – Starting up and adjustments 
UM-30300D-U User manual AX 6000 Page 22 / 54 
5. PARAMETERS ADJUST 
Now adjust the test cycle parameters.  
The process to adjust the parameters is the same for all.   
Press and hold the   key (three seconds) to enter into the 
main menu. 
 
 Parameters : Unlock
 Mode : Program 
>Program : Pr01 
 
By using the   key, move down up to the test parameters 
and validate with the 
 key, the cursor goes on the right 
side to confirm the parameter edition. 
 
 Name  : WING L 
 Current  : 10A 
>R Max : 1.000mΩ 
 
By using the   and   keys, adjust the parameter to 
the hope value and validate with the 
 key, the cursor back 
to the left side of the display. 
 
 Name  : WING L 
 Curreant  : 10A 
>R Max : 1.300m
Ω 
 
5.1. PARAMETERS TO ADJUST 
5.1.1.  Program name (NAME) 
This function allows personalizing a program, for example to identify it with the part to 
test name. Enter 7 alphanumeric characters maximum.  
5.1.2.  Measurement current (CURRENT) 
The instrument had three measurements current available, the value to use is based on 
the material of the parts to control:   
¾ 10 A for the metallic parts,  
¾ 1 A and 0.1 A for the composites material parts. 
5.1.3.  Maximum reject (R Max.) 
This parameter defines the threshold above the test part is considered as fail. If the 
measured value exceeds this value, the test is fail.