Chapter 3 – Starting up and adjustments
UM-30300D-U User manual AX 6000 Page 22 / 54
5. PARAMETERS ADJUST
Now adjust the test cycle parameters.
The process to adjust the parameters is the same for all.
Press and hold the key (three seconds) to enter into the
main menu.
MAIN
MENU
Parameters : Unlock
Mode : Program
>Program : Pr01
By using the key, move down up to the test parameters
and validate with the
key, the cursor goes on the right
side to confirm the parameter edition.
MAIN
MENU
Name : WING L
Current : 10A
>R Max : 1.000m
By using the and keys, adjust the parameter to
the hope value and validate with the
key, the cursor back
to the left side of the display.
MAIN
MENU
Name : WING L
Current : 10A
R Max : 1.300m
<
Adjust the others parameters by using the same process.
MAIN
MENU
Name : WING L
Curreant : 10A
>R Max : 1.300m
5.1. PARAMETERS TO ADJUST
5.1.1. Program name (NAME)
This function allows personalizing a program, for example to identify it with the part to
test name. Enter 7 alphanumeric characters maximum.
5.1.2. Measurement current (CURRENT)
The instrument had three measurements current available, the value to use is based on
the material of the parts to control:
¾ 10 A for the metallic parts,
¾ 1 A and 0.1 A for the composites material parts.
5.1.3. Maximum reject (R Max.)
This parameter defines the threshold above the test part is considered as fail. If the
measured value exceeds this value, the test is fail.