Chapter 6 –Error messages 
UM-30300D-U User manual AX 6000 Page 45 / 54 
 
Chapter 6 
 ERROR MESSAGES  
The ATEQ AX6000 can display error messages if there are operational problems. 
PROBLEM  LIGHTS  DISPLAYED MESSAGES 
O
O
V
V
E
E
R
R
 
 
R
R
A
A
N
N
G
G
E
E
 
 
The measure is out of the full scale.  
Action: take another measurement 
range, or configure the instrument in 
automatic range.  
 
  
 Program 1 
OVER 
RANGE 
10A  6mΩ
 
C
C
O
O
N
N
T
T
I
I
N
N
U
U
I
I
T
T
Y
Y
 
 
U
U
 
 
F
F
A
A
U
U
L
L
T
T
 
 
Contact break on the voltmeter loop. 
Action: restart a measurement, if the 
problem persists, check the contact 
quality accessories and fittings. 
 
 
 
 Program 1 
CONTINUITY U 
FAULT 
10A  Auto
 
C
C
O
O
N
N
T
T
I
I
N
N
U
U
I
I
T
T
Y
Y
 
 
I
I
 
 
F
F
A
A
U
U
L
L
T
T
 
 
Contact break on the current generator 
loop. 
Action: restart a measurement, if the 
problem persists, check the contact 
quality accessories and fittings. 
 
 
 
 Program 1 
CONTINUITY I 
FAULT 
10A  Auto
 
M
M
E
E
A
A
S
S
U
U
R
R
E
E
M
M
E
E
N
N
T
T
 
 
F
F
A
A
I
I
L
L
U
U
R
R
E
E
 
 
Measurement board failure.  
Action: turn off and restart the device, if 
the problem persists, contact the ATEQ 
after sale service. 
 
 
 
 Program 1 
MEASUREMENT 
FAILURE 
10A  Auto
 
A
A
U
U
T
T
O
O
T
T
E
E
S
S
T
T
 
 
A
A
L
L
A
A
R
R
M
M
 
 
Auto-test fault. The internal auto-test 
cycle result is fail, so the test current is 
on wrong value. 
Action: make one or several auto-test 
cycles on all the ranges, if the problem 
persists, contact the ATEQ after sale 
service. 
 
 
 
  AUTO-TEST 
102 mA 
 ALARM 
 
P
P
a
a
s
s
s
s
w
w
o
o
r
r
d
d
 
 
e
e
r
r
r
r
o
o
r
r
 
 
The password entered is wrong. 
Action: try again with the right 
password. 
 
 
Password Error