The maximum and minimum values measured are shown as an
expanded part of the scale (see Figure 4-52). To reset the maximum
and minimum values, press the BARGRAPH key. The printout from a
bargraph will also show the specific numerical values of maximum,
minimum, and current readings.
The center soft key allows the selection of AUTORANGE or
HOLDRANGE. In order to obtain the wide dynamic range necessary
for measuring audio levels, the amplitude meters are equipped with
ranging amplifiers, which automatically prescale the signal so that it is
in the best range for measurement. This is called autoranging. When
the ranging amplifiers change ranges, the reading may glitch. These
glitches can destroy the validity of the maximum and minimum
values. The autoranging may be turned off (holdrange mode) to
avoid these glitches, and the range will stay as currently selected. The
penalty to this is that there will be a maximum limit to the amplitude
that can be measured in this range, and the resolution may be limited
at low amplitudes. As long as the reading doesn’t vary much after
entering holdrange mode, the range will be ideal.
The speaker output is taken after the ranging amplifiers, so Holdrange
mode will also keep the speaker signal from autoranging. This may be
useful for audible monitoring of signals with a wide dynamic range.
Holdrange mode will be maintained after exiting bargraph mode, but
autorange will be selected if a new measurement function is chosen.
The units for the bargraph can be changed by pressing the upper right
soft key. This will change the units of both the numerical reading and
the bargraph. However, the units of the numerical reading may be
changed by going to the Main Panel and changing the units there
(using the same soft key). When you return to the Bargraph Panel, the
bargraph units will be the same as they were before, but the numerical
display will show the newly selected units. The ATS-1 Dual Domain
remembers separate units for each numerical display and the bargraph
for each measurement function.
4 Functions
Bargraphs Function Descriptions
4-52 ATS-1 Dual Domain User's Manual