6 Parameter Settings and Functions     
84    © Copyright Reserved Autonics Co., Ltd. 
6.4.6  Alarm output delay settings [PAR4 → A!ON / A!OF/ A@ON / A@OF/ 
A#ON / A#OF] 
Alarm output delay can be set to prevent false alarms caused by erroneous input signals 
resulting from disturbances or noise. 
With a preset delay time, alarm output does not turn on for the preset duration. Instead, the 
concerned alarm indicator on the front will flash in 0.5 sec. intervals. 
 
 
A!ON 
Alarm output 1 on delay: Stands by for the preset duration upon an alarm event, 
checks the alarm trigger conditions, and turns on the alarm output if the 
conditions are still present. 
A!OF 
Alarm output 1 off delay: Stands by for the preset duration following alarm output 
off, checks the alarm trigger conditions, and turns off the alarm output if the 
deactivation conditions are still present. 
A@ON 
Alarm output 2 on delay: Stands by for the preset duration upon an alarm event, 
checks the alarm trigger conditions, and turns on the alarm output if the 
conditions are still present. 
A@OF 
Alarm output 2 off delay: Stands by for the preset duration following alarm output 
off, checks the alarm trigger conditions, and turns off the alarm output if the 
deactivation conditions are still present. 
A#ON 
Alarm output 3 on delay: Stands by for the preset duration upon an alarm event, 
checks the alarm trigger conditions, and turns on the alarm output if the 
conditions are still present. 
A#OF 
Alarm output 3 off delay: Stands by for the preset duration following alarm output 
off, checks the alarm trigger conditions, and turns off the alarm output if the 
deactivation conditions are still present. 
 
Parameter  Setting range 
Unit 
PAR4 
A!ON 
0000 to 3600  0000  Sec. 
A!OF 
A@ON 
A@OF 
A@ON 
A@OF