64 Model E1070 Indicator User Instructions
Indicator Operations
Using Outputs in the
Counting Application
If enabled, you can use the output function while in the counting application.
Outputs are enabled in a password protected menu. Contact your local
Avery Weigh-Tronix supplier or distributor for assistance. Follow these
steps to set up to three outputs:
You can press the SELECT key or
the ESC key to abort the process in
step 3.
Output operation in the counting
application:
Below Configured Value:
Outputs are ON
Annunciators are OFF
TIU3 Relays are ON
Above Configured Value:
Outputs are OFF
Annunciators are ON
TIU3 Relays are OFF
1. With the indicator powered up, press and hold the SELECT key until…
OP1 is displayed.
2. Press the PRINT key…
The current value for OP1 is displayed.
3a). Press ENTER to accept this value
OR
3b). Key in a value and press the ENTER key.
OP1 is displayed.
4. Scroll through all three outputs by using the left and right arrow keys.
5. Repeat steps 2 and 3a or 3b for each output.
6. Press ESC key to exit the outputs setup.
7. As you apply weight to the scale, output one will activate above its
setpoint and deactivate below its setpoint. The same is true for the
other two setpoints. When activated each output’s annunciator will
light on the display.