A-8
Appendix A  Maintenance and Troubleshooting of this Device
A2-4-5  Setting of the diagnosis alarm
If diagnosis is judged to be possible, set the alarm. If you do not want to activate the alarm, omit the 
steps in this section.
[CAUTION]   The procedures described in this document do not guarantee the detection of 
clogging or the elimination of false indications. The pressure frequency index may 
change due to causes other than clogging, and may also depend on the degree of 
clogging or the material of the clog. Adjusting the settings should be done with the 
realization that there is no threshold value setting that can eliminate misinformation 
and securely detect clogging only.
First, determine the alarm threshold value. This threshold should be between the minimum index 
value under normal conditions and the maximum with simulated clogging. If the threshold is put 
close to the minimum under normal operating conditions, the alarm will be activated earlier, but 
the possibility of a false alarm is also higher. Put the threshold at a certain distance away from the 
minimum under normal operating conditions, leaving a margin.
Index value range 
under the normal 
operating conditions
Minimum value 
under normal 
operating conditions
Maximum value 
under simulated clogging
Threshold 
value
After determining the threshold value, follow the steps below to set the alarm and start the 
diagnosis.
Procedure
(1) Execute Reset Press Freq Index.
(2) Set Press Freq Index Low Limit to the determined threshold value.
(3) Set Press Freq Index Alarm Use “Enabled (Low)” (lower limit only).
With these settings, when the index value is outside the normal range and approaches the value 
with simulated clogging, the alarm is activated.