Azbil Corporation Operation using the data setting device
Model MTG11A/18A, MTG11B/18B, MTG14C 5-3
Display and operation contents of data setting device
Overview of mode
This system provides the following four modes available in accordance with the
operations:
Mode Description
MEASURING MODE Mode that shows measuring status.
OPERATOR’S MODE Mode that is set for the operator. This mode is comprised of
setting and configuration of data that are set or changed
frequently during startup. In this mode, settings can be
changed only when the write protect levels are set to 0, 1 and
2. At level 3, only the set data can be checked. (See “5-3-2 :
Display of write protect level”)
[Damping constant, auto zero adjustment, counter reset,
counter preset value, etc.]
m CAUTION
Set or changed data are temporarily written into the
memory. Note that if the configured data are not
saved/written into the memory within 10 minutes, the
configured data returns to the previous values.
Be sure to press the MODE key to return to the
MEASURING MODE and to save data.
ENGINEERING MODE In the set mode for engineering, the mode is comprised of data
that is set or changed less frequently than the data in
“OPERATOR'S MODE.”
Data can be set and changed at write protect level 0 or 1. At
level 2 or 3, only the set data check is allowed.
[ID, function selection, detector data, flow rate range,
hysteresis width, pulse data, low flow cut, selecting false mode
for output, etc.]
m CAUTION
When the mode is changed to MEASURING
MODE by pressing the MODE key, the set/changed
data are saved into a non-volatile memory. Be sure
to press the MODE key, to save the configured data.
Set or changed data are temporarily written into
the memory. Note that if the configured data are
not saved/written into the memory within 10
minutes, the configured data returns to the
previous values. Be sure to press the MODE key
to return to the MEASURING MODE and to
save data.