Azbil Corporation Instrument installation
Model MTG11A/18A, MTG11B/18B, MTG14C 2-13
Flange shape
The flanges used should be such that the area of contact with the gasket is maximized,
as shown in Figure 2-10.
Figure 2-10 Flange shape
~Note Before installing the detector be sure to flush out any foreign matter
that may be present inside the detector. Residual foreign matter could
cause output fluctuations.
Do not touch the electrodes by your hand or waste cloth with oil. It
could cause output fluctuations.
Install the detector in accordance with the flow direction mark on the
detector with the direction of the liquid flow. Misalignment could
result in a negative output.
Acceptable X Unacceptable
Flange
Welding
Welding
Pipe
(The liquid could
leak because of
the small area
of contact with
the gasket.)