EasyManua.ls Logo

Azbil MagneW Neo PLUS - Description of Error Messages

Azbil MagneW Neo PLUS
338 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Azbil Corporation Operation using the data setting device
Model MTG11A/18A, MTG11B/18B, MTG14C 5-77
5-8 : Description of Error Messages
Introduction
Errors are classified into critical failure and non-critical failure.
Critical failure
Critical failure may obstruct the electromagnetic flowmeter operation, if not corrected,
ultimately damage the flowmeter. When critical failure occurs during operation, an
error message will appear on the converters display and the electromagnetic
flowmeter will continue to output the preset value in the abnormality treatment
(failsafe) direction. The error message and the self-diagnostic results will be visible on
the display.
Perform the proper correction measures, referring to the actions below.
Error code of serious trouble
Error
code
Error content Action LCD display
Err-02 CPU (ROM, RAM)
CHECK SUM
ERROR
1. Restore power.
2. Replace ROM.
3. Replace main P/C.
Err-04 NVM
READ AFTER
WRITE ERROR
1. Restore power.
2. Replace main P/C.
Err - 02
CPU CHECK
ERROR
Err - 04
NVM CHECK
ERROR
CM2-MTG300-2001.book 77 ペー   

Table of Contents

Related product manuals