Operation using the data setting device Azbil Corporation
5-74 Model MTG11A/18A, MTG11B/18B, MTG14C
5-7-8 : Configuration of CRITICAL MODE
Introduction
CRITICAL MODE has the following setting and adjustment items:
LCD display flow
The screen flow of CRITICAL MODE is as follows:
Item Content Screen
ROM VER
DATE
Displays the ROM version and date.
SHIPPING DATA
(default value)
RECOVERY
You can return the device to factory
setting/default values for pertinent
operational and configuration
parameters.
These parameters are entered before
the device is shipped, so they are
commonly referred to as “shipping
data”.
They include factory calibration
data and factory settings or initial
default settings for customer
configuration data.
INITIAL
DATA
RECOVERY
Initial data recovery eliminates all
calibration data and configuration
parameters.
DO NOT use this function.
m CAUTION
INITIAL DATA RECOVERY
function is only for Azbil Corporation
service/maintenance specialist.
Please DO NOT use this function.
If this function is turned ON, all
calibrated data will be missing. The
device needs to be back to the factory
to calibrate again.
20.0 %
> ROM VER.
DATE YY-MM-DD
20.0 %
> SHIPPING DATA
RECOVERY READY
20.0 %
> INITIAL DATA
RECOVERY READY
> ROM VER. 3.00
DATE 02-07-15
> SHIPPING DATA
RECOVERY READY
> INITIAL DATA
RECOVERY READY
> OUTPUT CHECK
MODE OFF
CAUTION
CM2-MTG300-2001.book 74 ペー   

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Azbil MagneW Neo PLUS and is the answer not in the manual?

Azbil MagneW Neo PLUS Specifications

General IconGeneral
Measurement PrincipleElectromagnetic induction
Fluid Conductivity5 μS/cm or more
Ambient Temperature-20 to +60°C (-4 to +140°F)
Output Signal4-20 mA DC
Power Supply24 VDC or 100-240 VAC
CommunicationHART
Process ConnectionFlange
Enclosure RatingIP67
Operating Temperature-20 to +60°C (-4 to +140°F)
Protection RatingIP67
MaterialStainless steel
Repeatability±0.2% of reading