Operation using the data setting device Azbil Corporation
5-6 Model MTG11A/18A, MTG11B/18B, MTG14C
This section describes keys on the data setting device.
Name Description
MODE key
Enters OPERATOR'S MODE.
When parameters and configured data have been
changed in ENGINEERING MODE or
MAINTENANCE MODE, press this key to save the
data.
Right shift key
Moves the cursor to the right.
Decrement key
Changes the parameter at a cursor position.
Displays the previous screen.
If the key is pressed, when the
cursor is placed at the upper
left end (*, #, >), the screen
will change
If the key is pressed, when the
cursor is placed at a number,
the number is decremented.
If the cursor is placed at a
decimal point, the decimal
point moves rightward.
Increment key
Changes the parameter at a cursor position.
Displays the following screen.
If the key is pressed, when the
cursor is placed at the upper
left end (*, #, >), the screen
will change.
If the key is pressed, when the
cursor is placed at a number,
the number is incremented.
If the key is pressed, when the
cursor is placed at a decimal
point, the decimal point
moves to the left.
If the cursor is placed over
READY, pressing the key
starts operation.
* OPERATOR'S
MODE
Cursor
* DAMPING
001.0 S
Cursor
# 1.0000 m/s
SPAN 07. 069 m
3
/h
Cursor
* OPERATOR'S
MODE
Cursor
* DAMPING
001.0 S
Cursor
# 1.0000 m/s
SPAN 07. 069 m
3
/h
Cursor
* AUTO ZERO
READY
Cursor
CM2-MTG300-2001.book 6 ペー 01  