Operation using the data setting device Azbil Corporation
5-20 Model MTG11A/18A, MTG11B/18B, MTG14C
Set the moving average processing in accordance with the following procedure:
Step Procedure Screen
1 The screen at right is a screen display example in
MEASURING MODE (measurement state).
Press the MODE key.
2 The OPERATOR'S MODE screen appears for
approx. two seconds and then the damping setting
screen appears.
3 Press the or key to display the screen shown
on the right.
4 Press the key to move the cursor to the OFF
position.
5 Press the key to switch the OFF screen to the
ON screen.
Press the key to move the cursor to the value to
be changed.
6 Press the or key to display a value to be set.
7 Press the key to move the cursor back to the
position under *.
Press the MODE key to return to the
MEASURING MODE and save data.
m CAUTION
You have only ten minutes to return to MEASURING MODE to save the new value before
the system resets it to the previously saved value.
20.0 %
01.94 m
3
/h
WP0 00069401
20.0 %
* OPERATOR'S
MODE
20.0 %
*
DAMPING
005.0 s
20.0 %
*
AVERAGING
OFF
20.0 %
*
AVERAGING
OFF
20.0 %
*
AVERAGING
ON 01.0 s
20.0 %
*
AVERAGING
ON 05.0 s
20.0 %
*
AVERAGING
ON 05.0 s
CM2-MTG300-2001.book 20 ペー   