Azbil Corporation Operation using the data setting device
Model MTG11A/18A, MTG11B/18B, MTG14C 5-27
Step Procedure Screen
9
(Con-
tin-
ued)
(Branch 4)
When the result is (4) in Step 8
Check that the empty or scale on electrode status
is not detected when SENSITIVITY MID has
been set.
(Result)
If the empty or scale on electrode status is not
detected, use that SENSITIVITY MID setting
without change.
If the empty or scale on electrode status is
detected and the screen shown on the right
appears, this function is not available in that
environment. Set the function to OFF.
(Branch 5)
When the result is (5) in Step 8
Check that the empty or scale on electrode status
is not detected when SENSITIVITY HIGH has
been set.
(Result)
If the empty or scale on electrode status is not
detected, use that SENSITIVITY HIGH setting
without change.
If the empty or scale on electrode status is
detected and the screen shown on the right
appears, this function is not available in that
environment. Set the function to OFF.
0.0 %
EMPTY OR SCALE
ON ELECTRODE
0.0 %
EMPTY OR SCALE
ON ELECTRODE