Azbil Corporation Operation using the data setting device
Model MTG11A/18A, MTG11B/18B, MTG14C 5-29
Blink
"EMPTY OR SCALE
ON ELECTRODE"
NO
YES
Change the SENSITIVITY
to LOW when the flowtube is
filled with the process fluid.
Blink
"EMPTY OR SCALE
ON ELECTRODE"
NO
YES
Electrode Status Diagnostic
function is not applicable.
Change the SENSITIVITY
to OFF.
Use the Electrode Status
Diagnostic function with the
SENSITIVITY LOW.
Change the SENSITIVITY to LLL
when the pipe is empty or the scale
appears on electrode.
Blink
"EMPTY OR SCALE
ON ELECTRODE"
NO
YES
Change the SENSITIVITY to
LL when the flowtube is filled
with the process fluid.
Blink
"EMPTY OR SCALE
ON ELECTRODE"
YES
NO
Electrode Status Diagnostic
function is not applicable.
Change the SENSITIVITY
to OFF.
Blink
"EMPTY OR SCALE
ON ELECTRODE"
NO
YES
Electrode Status Diagnostic
function is not applicable.
Change the SENSITIVITY
to OFF.
Use the Electrode Status
Diagnostic function with the
SENSITIVITY LL.
Set the SENSITIVITY as
LLL when the flowtube is
filled with the process fluid.
Use the Electrode Status
Diagnostic function with the
SENSITIVITY LLL.