Azbil Corporation Operation using the data setting device
Model MTG11A/18A, MTG11B/18B, MTG14C 5-71
m CAUTION
When calibrate the remote style converterwith the MGZ calibrator, short C terminal and
ground terminal on the converter housing.
By shorting C terminal and ground terminal on the converter housing, accurate calibration
can be done.
Figure 5-4
<Remote style>
MGZ calibrator
SIGNAL
X
Y
B
A
C
Short C terminal and ground terminal
on the converter housing.
Grounding resistance
less than 100Ω
Grounding resistance
less than 100Ω
CM2-MTG300-2001.book 71 ページ 2015年9月29日 火曜日 午前10時14分