Azbil Corporation Operation using the data setting device
Model MTG11A/18A, MTG11B/18B, MTG14C 6-69
5 Display shown to the right appears.
Press the key and select the
value to be calibrated.
The value to gain calibrate for two
wired magnetic flowmeter are the
following three: 0 m/s, 2.5 m/s, and
10 m/s. Value 2.5 m/s for two
wired magnetic flowmeter will be
shown 1.2 m/s in SFC. When
calibrating to 2.5 m/s select 1.2 m/
s.
6 For example, when calibrating 2.5
m/s, displays screen as shown to
the right. For smart calibrator
model MGZ14, enter 2.5 m/s.
Then, press key. Gain
calibration starts. When gain
calibration completes, display will
return back to the screen to select
the gain calibration value.
7 Follow the same procedure for
calibrating 0 m/s and 10 m/s.
8 When gain calibration is
completed, press the key and
the display as shown on the right
appears.
9
Press the key.
Exits from gain calibration
function.
10
Press the key and the screen
returns to the screen as in step 1.
Step
Key Procedure SFC screen
MENU
I
ITEM
DE CONF.
MENU
I
ITEM
DE CONF.
SFCM00006125D
SFCM00006126D
ENTER
( Yes )
NON-VOL
ENTER
( Yes )
NON-VOL
SFCM00006126D
SFCM00006126D
SFCM00006128D
MENU
I
ITEM
DE CONF.
MENU
I
ITEM
DE CONF.
SFCM00006127D
ENTER
( Yes )
NON-VOL
ENTER
( Yes )
NON-VOL
SFCM00006128D
SFCM00006129D
CLR
( No )
CLR
( No )
SFCM00006009D
CM2-MTG300-2001.book 69 ペー   