Azbil Corporation Operation using the data setting device
Model MTG11A/18A, MTG11B/18B, MTG14C 6-77
5
Press the key and select the
pulse width to be set. (Select from
duty 50, 1/7/10/15/30/50/100/200
ms)
You can enter it directly by
pressing the numeric keys. Range
of two wired magnetic flowmeter
is 1 to 1000 ms, however, SFC can
only set up to 999.9 ms.
6
Press the key.
The changed setting is saved on the
SFC.
7
or
Press the key or key to
display the screen as shown on the
right.
8
Press the key.
Changed setting data is written into
the database of the converter.
9
Press the key and the screen
returns to the screen as in step 1.
Step
Key Procedure SFC screen
CM2-MTG300-2001.book 77 ページ 2015年9月29日 火曜日 午前10時14分