7 Troubleshooting Chart
7-22 FLO-GARD 6201 VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07-19-B1-688
The backlight cannot be turned off. N/A Replace Sensor board.
Replace Display board.
The pump turns off or stops with
audible alarm for no apparent reason.
Check ICP301.
Check CNBUS1/CNBUS2 for proper
connection and continuity.
Check front panel cables for short
circuits.
Repair or replace CNBUS1/CNBUS2
Replace front panel film.
Replace Sensor board.
Replace Display board.
ALARM LED is always lit. Turn pump off and back on. If problem recurs, charge or replace
battery.
Replace CPU board.
Replace Display board.
ALARM LED does not light. Check CNDISP and CNBUS1/
CNBUS2 for proper connection and
continuity.
Replace CPU board.
Replace Display board.
ALERT LED is always lit. Turn pump off and back on.
Check CNBUS1/CNBUS2 for proper
connection and continuity.
If problem recurs, charge or replace
battery.
Replace CPU board.
Replace Display board.
ALERT LED does not light. Check CNDISP for proper connection
and continuity.
Replace CPU board.
Replace Display board.
PUMPING LED is always lit. Check CNDISP for proper connection
and continuity.
Replace Sensor board.
Replace Display board.
PUMPING LED does not light. Check CNDISP for proper connection
and continuity.
Replace Sensor board.
Replace Display board.
The audible alarm volume is not loud
enough.
N/A Turn the VOLUME knob on the rear
of the pump clockwise until the
desired volume is obtained.
Audible alarm is always on. Momentarily disconnect and
reconnect battery.
Check F302 or ICP301. Check
CNBUS1/CNBUS2, CN302, CN701,
CN751, and CNI/F for proper
connection and continuity.
Charge or replace battery.
Replace Sensor board.
Replace CPU board.
Audible alarm does not sound. Turn pump off and back on. If problem recurs, replace CPU board.
Replace Sensor board.
Replace speaker.
Table 7-4 Troubleshooting Chart — continued
Symptom Checks Corrective Action