1 Overview of the Device
Infusomat® compact
plus
P 1.0 1 - 7
EN
For internal use, only
FUNCTIONAL TEST
*
Danger of injury to the patient!
Device must not be connected to a patient.
Observe the Instructions for Use when performing the test.
* Do not use the device if the functional test fails. Contact your local representative for replacement or repair.
Item OK Item OK
1. Completeness, Integrity 3. Functional Test (see Instructions for Use)
1.1 Instructions for Use is available. ❒ 3.1 Open pump door, insert a primed line, close pump
door.
❒
1.2 Final Inspection Sheet is available. ❒
1.3 No mechanical damage is found on Infusomat®
compact
plus
P.
❒ 3.2 Open roller clamp, confirm line. "Rate" menu appears
after successful self-test.
❒
1.4 Type plate is present. ❒ 3.3 Deny priming if this message appears (configurable). ❒
1.5 Read serial number from type plate and enter this
below on this page.
❒ 3.4 Select a rate of 1.1 ml/h and start Infusomat®
compact
plus
P: Green LED lights up, arrows move from
right to left.
❒
1.6 Seal is present and undamaged. ❒
1.7
Infusomat® compact
plus
P is clean.
❒ 3.5 Administer a bolus by preselecting a volume of 3 ml –
bolus is administered, volume counter counts up.
❒
2. Switching on and Self-test 3.6 Close roller clamp – red LED lights up, alarm text is
displayed.
❒
2.1
Switch on Infusomat® compact
plus
P.
❒
2.2 Display shows "Self test". ❒ 3.7 Confirm alarm by pressing the "OK" button and restart
Infusomat® compact
plus
P.
❒
2.3 Two signal tones are audible. ❒
2.4 Red LED flashes twice.
Green LED flashes once.
❒ 3.8 Press "Menu" button – the "Main Menu" opens. ❒
3.9 Press "Info" button – therapy information is given. ❒
2.5 Is the battery charged?
If not: connect device to mains power supply.
❒
4. Finalizing the Test
2.6 Is the required language displayed?
If not: select language.
❒ 4.1
Remove line from Infusomat® compact
plus
P and
dispose properly. Switch device off.
❒
4.2 Place Instructions for Use with Infusomat®
compact
plus
P.
❒
4.3 Complete this form and file properly. ❒
4.4 Also file the Final Inspection Sheet. ❒
Functional Test performed at Infusomat® compact
plus
P with serial number:
Infusomat® compact
plus
P was handed over without any damage and is in proper working order.
Date Signature B. Braun Signature Owner