4 Check after Repair
4 - 6 Infusomat® compact
plus
P 1.0
EN
For internal use, only
PROCEDURAL INSTRUCTIONS ON THE CHECK AFTER REPAIR T1 Visual inspection
1. Check the Infusomat® compact
plus
P for cleanliness,
completeness, damage, and defects affecting safety, damage,
and readability of the labels.
Pay special attention to the following parts:
a) Housing
Housing is clean and undamaged.
b) Screw covers
Screw covers on the housing present and undamaged.
c) Pump door
d) SHK ICPP
e) Air sensor
f) Pressure sensor, downstream
g) Membrane
Membrane in the housing front is undamaged
h) Pressure sensor, upstream
i) Second clamp
j) Keypad/display
Keypad/display is clean and undamaged, correct bonding of
the foil.
k) Accessory connector
Accessory connector is clean and undamaged.
l) IR window
IR window is clean and undamaged.
m) Mains power supply connection
Gasket for rubber connector CP (IEC socket) is present and
undamaged. Electrical contacts are clean and undamaged.
2. Check the accessories for cleanliness, completeness, damage,
and faults affecting safety, damage, and readability of the
labels.
Pay special attention to the following parts:
a) Mains power supply cable
b) Staff call cable compact
plus
c) Connection lead 12V for ambulance vehicles.
3. Check how the Infusomat® compact
plus
P and accessories are
configured.
14