Important Preliminary Remarks
Infusomat® compact
plus
P 1.0 0 - 9
EN
For internal use, only
MEASURING EQUIPMENT Additional measuring and test equipment used for servicing our
devices must comply with the following standards and/or mea
-
suring equipment specifications:
Electrical safety:
IEC 60601-1, ed 3.1
EN 62353
For example, Bender UNIMET 800/810ST
bender-de.com/en
or comparable
Voltage measurement: For example, Multimeter Fluke 177
fluke.com/en-us
or comparable
Clearance measurement:
Slip gauges
According to DIN 2275, tolerance T2 or better.
For example,
wermas.de
or comparable
Pressure measurements:
Measurement accuracy ≤ 80 mbar (1.16 psi)
For example, Kobold MAN-SD1S Y A3 0
S#2062378.8 customer-specific
kobold.com/en
or Sika MH 3750
sika.net
or Fluke 717
fluke.com
or comparable
Flow accuracy:
Scales: Measurement range 220 g
Linearity deviation ±0.1 mg
Standard deviation ±0.05 mg
Readability 0.01 mg